Z-1663-2010
Bausch & Lomb Millennium Microsurgical System
Bausch & Lomb Incorporated / initiated 2010-01-15 / Terminated / root cause: manufacturing process
Bausch & Lomb Inc began a recall of Bausch & Lomb Millennium Microsurgical System on . The reason FDA records is "The AC power cords used on these products are subject to an FDA Alert -- Safety Investigation of Certain Medical Device Power Cords: Initial Communication, Date Issued: October 19, 2009". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1663-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bausch & Lomb Incorporated
- Product
- Bausch & Lomb Millennium Microsurgical System.
- Product code
- NFJ System, Image Management, Ophthalmic
- Reason for recall
- The AC power cords used on these products are subject to an FDA Alert -- Safety Investigation of Certain Medical Device Power Cords: Initial Communication, Date Issued: October 19, 2009.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- A "URGENT: POWER CORD RECALL" letter and copies of the FDA Alert dated Oct. 19, 2009 were sent to all customers on or about Jan. 15, 2010, by Stericycle, a recall notification service. The letter describes the product, problem and actions to be taken by the customers as well as FDA's recommendations and actions. The customers were made aware of the FDA Alert concerning the AC power cords and notified that these power cords were used in the devices. They were requested to inspect the power cords for damage and notify the recalling firm if damage is noted or if they have experienced sparking, charring, or firing from the power cords and/or cracking or failure of the prongs. In the absence of those conditions, customers were given the option of returning the power cord for replacement and/or during the regularly scheduled servicing of the device. If you would like a new power cord, please contact Customer Service at 1-800-338-2020 for a free replacement.
- Quantity in commerce
- Quantity never provided by firm
- Distribution
- Worldwide distribution: United States & Puerto Rico; Venezuela, India, Trinidad and Tobago, Colombia, Sri Lanka, Hong Kong, Vietnam, and Peru.
- Product codes and lots
- All devices manufactured since 2004