Journey " Uni Tibial Baseplate, Right Medial/left Lateral, A/p 42 Mm, M/l 25 Mm, Ref 71422232, for Cemented...
Smith & Nephew Inc began a recall of JOURNEY " UNI TIBIAL BASEPLATE, RIGHT MEDIAL/LEFT LATERAL, A/P 42 MM, M/L 25 MM, REF 71422232, FOR CEMENTED USE ONLY, QTY: (1), STERILE R, SMITH & NEPHEW, INC... on . The reason FDA records is "The firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0930-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Journey " Uni Tibial Baseplate, Right Medial/left Lateral, A/p 42 Mm, M/l 25 Mm, Ref 71422232, for Cemented Use Only, Qty: (1), Sterile R, Smith & Nephew, Inc. Memphis, Tn 38116
- Product code
- HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
- Reason for recall
- The firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 12/23/2009. All affected Smith & Nephew International Distributors were notified via e-mail and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. For questions, please contact the firm, Leah Easley at 901-399-6137.
- Quantity in commerce
- 628 units
- Distribution
- US and Argentina, Switzerland, Australia, Canada, Belgium, Italy, Spain, Thailand, Austria; PROC, Finland, Germany, Malaysia, Portugal, Netherlands, Singapore, and England
- Product codes and lots
- All Lot Numbers