Z-1790-2010
Triage Total Controls Calibration Verification 5: Two - 0.25 mL tubes for each of five different levels...
Alere San Diego, Inc. / initiated 2009-02-06 / Terminated
Biosite Inc began a recall of Triage Total Controls Calibration Verification 5: Two - 0.25 mL tubes for each of five different levels (A,B,C,D,E). Storage Temp <-20C. Intended use... on . The reason FDA records is "Incidence of low recovery results has exceeded post market surveillance thresholds of customer complaints in the field". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1790-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Triage Total Controls Calibration Verification 5: Two - 0.25 mL tubes for each of five different levels (A,B,C,D,E). Storage Temp <-20C. Intended use: Materials used with the Triage Profiler S.O.B. Panel, Triage CardioProfilER Panel, Triage BNP Test, Triage D-Dimer Test and Triage Meters to verify the calibration of the Test Devices throughout the measurable range.
- Product code
- jjy
- Reason for recall
- Incidence of low recovery results has exceeded post market surveillance thresholds of customer complaints in the field.
- FDA root cause
- Other
- Action
- The firm, Biosite Incorporated, sent an "URGENT MEDICAL DEVICE RECALL" dated February 6, 2009, to all customers. The letter describes the product, problem and action to be taken by the customers. The customers were instructed to immediately discontinue all use and/or sale of the listed products, discard all affected product in accordance with their local regulations, complete and fax the enclosed Verification Form within 10 days to confirm their receipt of the notice, to indicate the number of kits discarded from their inventory and the product replacement will be sent to them. Contact Inverness Medical Technical Service at 1-877-441-7440 and select option 2 from the phone prompt and a representative will confirm the number of kits discarded from your inventory and arrange for replacement product to be sent to you. Should you have any questions about the information contained in this notification, please contact (877) 441-7440 or email: [email protected]
- Distribution
- Nationwide distribution
- Product codes and lots
- Model 88755 Lots 194444 & 194450