Z-1234-2010

ATAC Serum Control Kit. The device is used for monitoring the performance of clinical chemistry methods...

ELITechGroup Inc / initiated 2009-12-04 / Terminated

Vital Diagnostics, Inc. began a recall of ATAC Serum Control Kit. The device is used for monitoring the performance of clinical chemistry methods within the clinical range on . The reason FDA records is "Vital Diagnostics observed a bias in the Quality Control Consensus Program between published values for ALP on the ATAC 8000 and reported customer means. The ranges were adjusted to reflect the analyzer population". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1234-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ELITechGroup Inc
Product
ATAC Serum Control Kit. The device is used for monitoring the performance of clinical chemistry methods within the clinical range.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Vital Diagnostics observed a bias in the Quality Control Consensus Program between published values for ALP on the ATAC 8000 and reported customer means. The ranges were adjusted to reflect the analyzer population.
FDA root cause
Other
Action
Customers were mailed copies of Technical Bulletin 8100 and 5024 dated September 30, 2009 which advised of the Alkaline Phosphatase QC value revisions for the Envoy 500 and the ATAC 8000. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822.
Quantity in commerce
802 Kits
Distribution
Nationwide.
Product codes and lots
Level 1, Lot Number 9137010 and Level 2, Lot Number 9138010.