Z-1135-2010
Philips Switched Internal Defibrillator Paddles (used with Philips HeartStart XL and HeartStart MRx...
Philips North America LLC / initiated 2010-01-19 / Terminated / root cause: manufacturing process
Philips Healthcare Inc. began a recall of Philips Switched Internal Defibrillator Paddles (used with Philips HeartStart XL and HeartStart MRx monitor/defibrillators) used during open heart surgery... on . The reason FDA records is "Paddle shock switch may fail to actuate and delay or prevent delivery of defibrillation therapy". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1135-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Switched Internal Defibrillator Paddles (used with Philips HeartStart XL and HeartStart MRx monitor/defibrillators) used during open heart surgery. Models: M4741A - Extra Large Switched internal Paddles M4742A - Large Switched internal Paddles M4743A - Medium Switched internal Paddles M4744A - Small Switched internal Paddles
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Paddle shock switch may fail to actuate and delay or prevent delivery of defibrillation therapy
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Philips issued Urgent Medical Device Recall notification to customers by letter on 1/25/10 via UPS advising of this defect and providing instructions on how to determine if they have a defective paddle. The customers will be assisted by field service engineers to remove and replace these defective units (those within the date range of September 1, 2008 to December 30, 2008). Customers are asked to follow the Actions to be taken by customer/user section of the Urgent Medical Device Recall Notification. Removed units will be replaced free of charge and will be destroyed by Philips. Distribution of the Urgent Medical Device Recall Notification will be managed by Philips representatives in each affected geography. Customers are asked to follow the Actions to be taken by customer/user section of the Urgent Medical Device Recall Notification.
- Quantity in commerce
- 1592 units
- Distribution
- Worldwide distribution: USA, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, Estonia, Finland, France, Georgia, Germany, HONG KONG, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Lebanon, Lithuania, Malaysia, Mauritius, Netherlands, NICARAGUA, Norway, Peru, Philippines, POLAND, Portugal, Puerto Rico, Russian Federation, Saudi Arabia, Serbia & Montenegro, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Turkey, UKRAINE, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, and Vietnam.
- Product codes and lots
- Units with date codes (mmyy) of 0908, 1008, 1108, 1208 (Units manufactured by Philips between September 1, 2008 and December 30, 2008)