CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St...
Smiths Medical ASD, Inc. began a recall of CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St. Paul, MN 55112 on . The reason FDA records is "The CADD-Legacy 1 Model 6400 Ambulatory Infusion Pump (Serial Number 370969) was shipped without documentation that the required accuracy testing had been performed during the manufacturing process. If the subject...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1235-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St. Paul, MN 55112.
- Product code
- FRN Pump, Infusion
- Reason for recall
- The CADD-Legacy 1 Model 6400 Ambulatory Infusion Pump (Serial Number 370969) was shipped without documentation that the required accuracy testing had been performed during the manufacturing process. If the subject infusion pump did not go through the accuracy testing, and if it does not meet accuracy specifications, there is a potential for over or under-delivery.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Smiths Medical telephoned the customer on December 23, 2009, to return the CADD-Legacy 1 Ambulatory Infusion Pump, serial number 370969, since there was no record of the completed accuracy testing for the infusion pump.
- Quantity in commerce
- 1
- Distribution
- MI
- Product codes and lots
- Serial number: 370969