Z-0807-2010

StatSpin Express 4 Horizontal Centrifuge Model # M510

StatSpin, Inc., dba Iris Sample Processing / initiated 2010-01-27 / Terminated / root cause: design

StatSpin, Inc., dba Iris Sample Processing began a recall of StatSpin Express 4 Horizontal Centrifuge Model # M510. Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes on . The reason FDA records is "Centrifuge shield micro-switch failed due to damage, and the unit opened, ejecting pieces". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0807-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
StatSpin, Inc., dba Iris Sample Processing
Product
StatSpin Express 4 Horizontal Centrifuge Model # M510. Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes.
Product code
JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
Reason for recall
Centrifuge shield micro-switch failed due to damage, and the unit opened, ejecting pieces.
Root cause tag
design
FDA root cause
Device Design
Action
Iris Sample Processing notified distributors by an Urgent Device Recall letter advising of the recall and the defect. Users are to immediately quarantine the affected products and choose between the two options listed in regards to correcting the affected centrifuge. Customers may return the unit or an upgrade kit with instructions will be sent to upgrade the unit. Questions should be directed to Gail Magrath at 800-782-8774 ext. 6123.
Quantity in commerce
1510 units
Distribution
Worldwide Distribution -- USA, CANADA, AUSTRALIA, JORDAN, UK, BELGIUM, FRANCE, CZECH, RUSSIA, KUWAIT', NORWAY, GREECE, TURKEY, DENMARK, FINLAND, SEOUL, KOREA, and HONG KONG.
Product codes and lots
Serial Numbers: 00100 through 001679.