Medtronic Orthopedic Dovetail Clamp, REF 9730289, Medtronic Navigation, Louisville, CO 80027. In combination...
Medtronic Navigation, Inc began a recall of Medtronic Orthopedic Dovetail Clamp, REF 9730289, Medtronic Navigation, Louisville, CO 80027. In combination with a tera-tracker reference frame, is used by an... on . The reason FDA records is "Orthopedic clamp may slip if used with non-indicated system, resulting in possible inaccurate navigation information being displayed". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0923-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic Orthopedic Dovetail Clamp, REF 9730289, Medtronic Navigation, Louisville, CO 80027. In combination with a tera-tracker reference frame, is used by an Orthopedic surgeon to navigate the location of an acetabular cup positioner/impactor during hip replacement surgery.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Orthopedic clamp may slip if used with non-indicated system, resulting in possible inaccurate navigation information being displayed.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- All affected users were notified by a Product Correction Notification letter on 11/03/2009 providing a description of the issue and informing them of the updated instructions for use for the Orthopedic Dovetail Clamp. Customers are to ensure that all users read the letter, place a copy of the letter with the IFUs for the related software applications listed in the letter, and fill out the attached response form indicating the receipt of the information. Further information is available at 720-890-3309.
- Quantity in commerce
- 244 units
- Distribution
- Worldwide Distribution -- USA, including VA, IA, MN and OK and countries of Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Hong Kong, Iceland, India, Israel, Italy, Japan, Korea, Lebanon, Malta, Mexico, Netherlands, Poland, Russian Federation, Saudi Arabia, South Africa, Switzerland, Turkey, United Arab Emirates, and UK.
- Product codes and lots
- All lot numbers.