Z-1138-2010

Modified Novy Cornual Cannulation Set with clear tip, G17478, REF J-NCS-503570, Cook OB/GYN

Cook Women'S Health / initiated 2010-01-08 / Terminated / root cause: manufacturing process

Cook Women'S Health began a recall of Modified Novy Cornual Cannulation Set with clear tip, G17478, REF J-NCS-503570, Cook OB/GYN on . The reason FDA records is "The distal tip may separate from the catheter while the catheter is in the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1138-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Women'S Health
Product
Modified Novy Cornual Cannulation Set with clear tip, G17478, REF J-NCS-503570, Cook OB/GYN.
Product code
MOV Catheters, Salpingography
Reason for recall
The distal tip may separate from the catheter while the catheter is in the patient.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Cook Women's Health sent Urgent Recall Notice letters dated 1/8/2010. Consignees were instructed to return product to the manufacturer for credit.
Quantity in commerce
2875
Distribution
USA, Ireland, China, Canada, India, Australia, Hong Kong.
Product codes and lots
U1705449, U1707347, U1709250, U1711011, U1714331, U1734602, U1737151, U1743919, U1746659, U1746660, U1755767, U1772781, U1772782, U1791573, U1792176, U1793188, U1799467, U1825725, U1825726, U1841365, U1862710 and U1863536.