Z-1138-2010
Modified Novy Cornual Cannulation Set with clear tip, G17478, REF J-NCS-503570, Cook OB/GYN
Cook Women'S Health / initiated 2010-01-08 / Terminated / root cause: manufacturing process
Cook Women'S Health began a recall of Modified Novy Cornual Cannulation Set with clear tip, G17478, REF J-NCS-503570, Cook OB/GYN on . The reason FDA records is "The distal tip may separate from the catheter while the catheter is in the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1138-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Women'S Health
- Product
- Modified Novy Cornual Cannulation Set with clear tip, G17478, REF J-NCS-503570, Cook OB/GYN.
- Product code
- MOV Catheters, Salpingography
- Reason for recall
- The distal tip may separate from the catheter while the catheter is in the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cook Women's Health sent Urgent Recall Notice letters dated 1/8/2010. Consignees were instructed to return product to the manufacturer for credit.
- Quantity in commerce
- 2875
- Distribution
- USA, Ireland, China, Canada, India, Australia, Hong Kong.
- Product codes and lots
- U1705449, U1707347, U1709250, U1711011, U1714331, U1734602, U1737151, U1743919, U1746659, U1746660, U1755767, U1772781, U1772782, U1791573, U1792176, U1793188, U1799467, U1825725, U1825726, U1841365, U1862710 and U1863536.