Dimension IBCT Flex Reagent Cartridge. The total (IBCT) method used on the Dimension system is an in vitro...
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension IBCT Flex Reagent Cartridge. The total (IBCT) method used on the Dimension system is an in vitro diagnostic test intended to quantitatively measure... on . The reason FDA records is "Test produces falsely elevated IBCT results and Abnormal Reaction test report messages on heparinized plasma samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2385-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension IBCT Flex Reagent Cartridge. The total (IBCT) method used on the Dimension system is an in vitro diagnostic test intended to quantitatively measure Total Iron Binding capacity in human serum and plasma.
- Product code
- JMO Ferrozine (Colorimetric) Iron Binding Capacity
- Reason for recall
- Test produces falsely elevated IBCT results and Abnormal Reaction test report messages on heparinized plasma samples.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens issued an "Urgent Field Safety Notice" letter dated December 2009 to users identifying the issue and affected product. Customers were instructed not to use heparinized plasma samples for IBCT testing. IBCT Instructions for Use (IFU) has been updated to reflect that heparinized plasma is no longer an acceptable sample type. Beginning with lot ED0341, the new IFU will be packaged in the carton with the IBCT Flex reagent cartridges. Consignees were requested to forward the notification letter to anyone to whom the product was distributed. The Siemens Technical Solutions Center can be contacted at 800-441-9250.
- Quantity in commerce
- unknown - all lots involved
- Distribution
- Worldwide Distribution: USA (all states) and the countries of Canada, Switzerland, New Zealand, Australia, Japan, and South Korea.
- Product codes and lots
- Catalog number df84