Dimension IBCT Flex Reagent Cartridge. The total (IBCT) method used on the Dimension system is an in vitro...

Siemens Healthcare Diagnostics Inc. / initiated 2009-12-16 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension IBCT Flex Reagent Cartridge. The total (IBCT) method used on the Dimension system is an in vitro diagnostic test intended to quantitatively measure... on . The reason FDA records is "Test produces falsely elevated IBCT results and Abnormal Reaction test report messages on heparinized plasma samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2385-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension IBCT Flex Reagent Cartridge. The total (IBCT) method used on the Dimension system is an in vitro diagnostic test intended to quantitatively measure Total Iron Binding capacity in human serum and plasma.
Product code
JMO Ferrozine (Colorimetric) Iron Binding Capacity
Reason for recall
Test produces falsely elevated IBCT results and Abnormal Reaction test report messages on heparinized plasma samples.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens issued an "Urgent Field Safety Notice" letter dated December 2009 to users identifying the issue and affected product. Customers were instructed not to use heparinized plasma samples for IBCT testing. IBCT Instructions for Use (IFU) has been updated to reflect that heparinized plasma is no longer an acceptable sample type. Beginning with lot ED0341, the new IFU will be packaged in the carton with the IBCT Flex reagent cartridges. Consignees were requested to forward the notification letter to anyone to whom the product was distributed. The Siemens Technical Solutions Center can be contacted at 800-441-9250.
Quantity in commerce
unknown - all lots involved
Distribution
Worldwide Distribution: USA (all states) and the countries of Canada, Switzerland, New Zealand, Australia, Japan, and South Korea.
Product codes and lots
Catalog number df84