Z-2287-2010
Deknatel Bondek Plus Polyglycolic Absorbable Surgical Sutures, Bon+VI R 0 HRX27 1N (70 cm), Teleflex Medical
Teleflex Medical / initiated 2010-02-01 / Terminated / root cause: labeling and UDI
Telefelx Medical began a recall of Deknatel Bondek Plus Polyglycolic Absorbable Surgical Sutures, Bon+VI R 0 HRX27 1N (70 cm), Teleflex Medical on . The reason FDA records is "Labeling error; The Bondek Suture product inside the box are of a different catalog number than labeled on the outer box". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2287-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Deknatel Bondek Plus Polyglycolic Absorbable Surgical Sutures, Bon+VI R 0 HRX27 1N (70 cm), Teleflex Medical.
- Product code
- GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
- Reason for recall
- Labeling error; The Bondek Suture product inside the box are of a different catalog number than labeled on the outer box.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Teleflex Medical issued consignees a "Urgent Medical Device Recall" letter dated February 1, 2010. Distributors were instructed to forward the letter to their consignees to retrieve relevant product from those locations. The letter described the packaging issue and potential impact. Customers were instructed to immediately discontinue use and quarantine any identified products, complete and fax the Recall Acknowledgement Form to 866-804-9881, in order to receive a Return Goods Authorization (RGA) Number, and then return the product to Teleflex Medical for replacement. Affected products received by Teleflex Medical will be destroyed upon receipt.
- Quantity in commerce
- 17 sales units (24 packs per sales unit/408 units)
- Distribution
- Worldwide Distribution: USA and Germany
- Product codes and lots
- Catalog Number: EP1081P, Lot number: 02A0801448, 02A0802800, 020A802802 and 02A0801433.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
