Medtronic N'Vision 8870 Application Card for use in the Medtronic model 8840 Physician Programmer...
Medtronic Neuromodulation began a recall of Medtronic N'Vision 8870 Application Card for use in the Medtronic model 8840 Physician Programmer. Manufactured for: Medtronic, Inc., Minneapolis, MN... on . The reason FDA records is "Routine monitoring of Medtronic's vigilance program identified user related programming errors. Medtronic has received reports of both overdose and underdose that were the result of programming use of SynchroMed...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1214-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic N'Vision 8870 Application Card for use in the Medtronic model 8840 Physician Programmer. Manufactured for: Medtronic, Inc., Minneapolis, MN 55432-5604 USA.
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Routine monitoring of Medtronic's vigilance program identified user related programming errors. Medtronic has received reports of both overdose and underdose that were the result of programming use of SynchroMed implantable infusion pumps. These errors were most commonly associated with: --Priming Bolus Programming; and --Bridge Bolus Programming. Pump programming errors that result in overdos
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Medtronic is exchanging 8870 versions AAG, AAJ and BBG and replacing it with the updated versions AAH01 and BBH01. The updated versions contain user interface improvements to reduce the chance of user related programming errors when programming SynchroMed II and SynchroMed EL Infusion Pump system. A Medtronic "Medical Device Correction" letter was left behind for consignees. The letter was addressed to "Healthcare Provider" and was dated February 2009. The letter provided information regarding changes to the programming software used in SynchorMed Infusion System. For additional information, please contact your Medtronic Representative or Medtronic Neuromodulation Technical Services at 1-800-707-0933.
- Quantity in commerce
- 8499 US
- Distribution
- All States in the US including DC. OUS: Austria Belgium Croatia Cyprus Czech Republic Denmark Estonia Faroe Islands Finland France French Polynesia Germany Greece Hungary India Ireland Israel Italy Kuwait Lebanon Malta Martinique Netherlands New Caledonia Norway Poland Portugal Reunion Russian Federation San Marino Saudi Arabia South Africa Spain Sweden Switzerland Turkey United Kingdom Vatican City State Canada Mexico Brasil
- Product codes and lots
- All model 8870 application cards version AAG, AAJ (English only US & Japan) & BBG (Multi-Language OUS)