Z-0806-2010

Endotek Alimaxx-B Uncovered Biliary Stent for Endoscopic Delivery, REF 70129-210, 10 x 80, Sterile EO, Merit...

Merit Medical Systems, Inc. / initiated 2010-01-28 / Terminated

Merit Medical Systems, Inc. began a recall of Endotek Alimaxx-B Uncovered Biliary Stent for Endoscopic Delivery, REF 70129-210, 10 x 80, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095... on . The reason FDA records is "Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0806-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Endotek Alimaxx-B Uncovered Biliary Stent for Endoscopic Delivery, REF 70129-210, 10 x 80, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Intended for palliation of malignant neoplasms in the biliary tree.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.
FDA root cause
Component change control
Action
Merit Medical Systems, Inc initiated customer notification by phone and letter beginning January 28, 2010. Consignees were instructed to immediately contact their affected customers, advise them of the recall activity and to identify and quarantine any unused inventory. A Recall Notification Form is to be completed and signed by the sales rep and by the site representative. Affected product is to returned to the firm. For further information, contact Merit Medical Systems, Inc. at 1-801-208-4228.
Quantity in commerce
51 units
Distribution
Nationwide Distribution -- AL, CA, CO, FL, IL, KY, NY, MA, MD, ME, MN, MO, PA, TX, UT, VA, WI, VA, TX and WI.
Product codes and lots
Lot Numbers: F727583A, IML4453A, IML4487A, IML4488A and IML4533A.