Z-0862-2010

Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, REF 011901, Sterile EO, Assembled In Mexico...

Bard Access Systems, Inc. / initiated 2010-01-14 / Terminated

Bard Access Systems began a recall of Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, REF 011901, Sterile EO, Assembled In Mexico, Bard Access Systems, Inc., SLC, UT 84116. Safety IV... on . The reason FDA records is "After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0862-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, REF 011901, Sterile EO, Assembled In Mexico, Bard Access Systems, Inc., SLC, UT 84116. Safety IV administration set used to access surgically implanted vascular ports.
Product code
FPA Set, Administration, Intravascular
Reason for recall
After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury.
FDA root cause
Component change control
Action
Bard Access Systems, Inc. issued and "Urgent: Huber Plus Non-Coring Needle Safety Infusion Set Recall" notification beginning January 2010. Users were instructed to quarantine all affected product and to notify any of their customers that had received the product. Quarantined product is to be returned to the firm. For further information, contact Bard Access Systems, Inc. at 1-800-290-1689.
Quantity in commerce
775 units
Distribution
Worldwide Distribution -- United States (VA, KY, MA, NJ and TN) and Belgium.
Product codes and lots
Lot Number: RETJ0656.