SpeedLink 6.0 Transverse Connector X-Long 60-75mm, non-sterile, 1304-60XL
Zimmer Inc. began a recall of SpeedLink 6.0 Transverse Connector X-Long 60-75mm, non-sterile, 1304-60XL on . The reason FDA records is "The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. The firm added...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0033-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- SpeedLink 6.0 Transverse Connector X-Long 60-75mm, non-sterile, 1304-60XL
- Product code
- KWP Appliance, Fixation, Spinal Interlaminal
- Reason for recall
- The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. The firm added additional instructions for use to reduce the probability of these two types of events.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The firm provided a revised surgical technique document to the Zimmer spine sales force in January, 2009. The firm will provide an Urgent Medical Device Correction notification to speedlink users through the sales force.
- Quantity in commerce
- 23,205 Units for ALL PRODUCTS IN THIS RECALL
- Distribution
- The devices are contained in implant cases stocked by Zimmer spine and made available to the Zimmer sales force for surgeries. They are not in general distribution.
- Product codes and lots
- 12LZ, 14MN, 36FB, 40EH and 8BR.