Z-2224-2010
Fresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CA
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2009-11-20 / Terminated / root cause: design
Fresenius Medical Care North America began a recall of Fresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CA on . The reason FDA records is "Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2224-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Fresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CA
- Product code
- FKP System, Dialysate Delivery, Single Patient
- Reason for recall
- Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- An urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the cord or plug.
- Quantity in commerce
- 33
- Distribution
- Product distributed to 5404 customers throughout the US and the world.
- Product codes and lots
- Catalog numbers 501228, R190188 (Refurbished), R501228, RTLR501228 (Refurbished)