Z-1551-2010

Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia...

Leica Microsystems, Inc. / initiated 2010-02-09 / Terminated

Leica Microsystems, Inc. began a recall of Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0001 and 26.0005 on . The reason FDA records is "Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve leakage have contributed to tissue damage during processing, leading to rebiopsies". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1551-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0001 and 26.0005
Product code
IEO Processor, Tissue, Automated
Reason for recall
Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve leakage have contributed to tissue damage during processing, leading to rebiopsies.
FDA root cause
Other
Action
Leica Microsystems sent Medical Device Correction letters dated 2/9/10 to the affected customers via first class mail on the same date, informing them of the potential for tissue damage during processing and outlining the software upgrade being made to minimize the risk of tissue damage. Customers were informed that their local Leica Microsystems representative will contact them to arrange the software upgrade and to provide training on the use of the new software. The letters also provided the necessary steps required to avoid tissue damage during processing while they are awaiting the software upgrade. The accounts were asked to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.
Quantity in commerce
574 units
Distribution
Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia.
Product codes and lots
catalog number 26.0001 and 26.0005, serial numbers 0260001B to 0260341B, 0260344B, 0260346B to 0260353B, and 0260356B to 0260366B.