Femoral Sponge 10 inches REF 206-715, Rx Sterile, Stryker Instruments Kalamazoo, MI. The femoral sponge is an...
Stryker Instruments Div. of Stryker Corporation began a recall of Femoral Sponge 10 inches REF 206-715, Rx Sterile, Stryker Instruments Kalamazoo, MI. The femoral sponge is an absorbent that is used to dry the intramedullary... on . The reason FDA records is "The firm did not have packaging validation to support the labeled shelf life. There is potential for a breach in sterility specific to the sterile packaged femoral sponge which may result in a non-sterile product". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1104-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Femoral Sponge 10 inches REF 206-715, Rx Sterile, Stryker Instruments Kalamazoo, MI. The femoral sponge is an absorbent that is used to dry the intramedullary canal prior to the application of bone cement. The suction effectively absorbs blood and fluid from the canal. It may provide swabbing/scrubbing action during suction.
- Product code
- GDY Gauze/Sponge, Internal, X-Ray Detectable
- Reason for recall
- The firm did not have packaging validation to support the labeled shelf life. There is potential for a breach in sterility specific to the sterile packaged femoral sponge which may result in a non-sterile product.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- A "Medical Device Recall Notification" dated January 21, 2010 was issued to customers. The notification described the affected product, issue and actions to take. Consignees were asked to forward the notification to all their affected locations. Customers were instructed to immediately check inventory and quarantine any recalled product found. Consignees were instructed to destroy all product and report how much product was destroyed. For questions regarding this recall, please contact Stryker Instruments by calling 1-800-800-4236, ext. 4354
- Quantity in commerce
- 9,257 all products
- Distribution
- Worldwide Distribution -- United States, Canada, Mexico, Australia, Brazil, Switzerland, Japan, Netherlands, England, UK, Sweden, Hong Kong, Korea, Malaysia, Singapore, Argentina, Spain, and Italy.
- Product codes and lots
- 743288, 743289, 743330, 743331, 743332, 743333, 743362, 743363, 743364, 743365, 743378, 743379, 743380, 743381, 743390, 743391, 743400, 743402, 743403, 743404, 743405, 743418, 743419, 743420, 743421, 743422, 743423, 743445, 743446, 743447, 743448 and 743449.