Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2356-2010

Stryker Pain Pump 2 Luer Cap, REF 525-401, Rx, Sterile, Product of Ireland Manufactured for Stryker...

Stryker Corp. / initiated 2010-02-01 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker Pain Pump 2 Luer Cap, REF 525-401, Rx, Sterile, Product of Ireland Manufactured for Stryker Instruments Kalamazoo MI on . The reason FDA records is "Defects include, but not limited to: Not sterile, not properly labeled, drugs and swabs may be ineffective, and catheter tips may break in patients". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2356-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker Pain Pump 2 Luer Cap, REF 525-401, Rx, Sterile, Product of Ireland Manufactured for Stryker Instruments Kalamazoo MI
Product code
FRN Pump, Infusion
Reason for recall
Defects include, but not limited to: Not sterile, not properly labeled, drugs and swabs may be ineffective, and catheter tips may break in patients.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker Instruments issued an "Urgent Medical Device Recall Notification" letter to consignee risk managers, directors and material managers on February 2, 2010, identifying the affected devices and product issues. The consignees were instructed to: 1) Forward the recall notice and reply form to all affected locations. 2) Review all inventory and quarantine affected product. 3) Destroy all affected product using attached instructions. 4) Complete and return the Business Reply Form to the firm by fax. Stryker Instruments can be contacted at 1-9800-800-4236 ext 4354
Quantity in commerce
18,826 all products
Distribution
Nationwide Distribution: United States, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, and UT.
Product codes and lots
2005060903, 2005070401.