Z-1421-2010

Cordis, AVANTI+ Introducer, STERILE EO, For one use only. Catalog 402656X Lot # 15076369. Catalog 504606X Lot...

Cordis US Corp. / initiated 2010-01-28 / Terminated / root cause: packaging

Cordis Corporation began a recall of Cordis, AVANTI+ Introducer, STERILE EO, For one use only. Catalog 402656X Lot # 15076369. Catalog 504606X Lot # 15075717. Catalog 504656X Lot # 15076362... on . The reason FDA records is "Sterility compromised: Cordis found a pin hole in one of the packaging tray cavities.On January 28, 2010 Cordis Corporation initiated a recall on their AVANTI + Introducer. During routine inspection, a visual...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1421-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cordis US Corp.
Product
Cordis, AVANTI+ Introducer, STERILE EO, For one use only. Catalog 402656X Lot # 15076369. Catalog 504606X Lot # 15075717. Catalog 504656X Lot # 15076362, 15076363, 15076364.
Product code
DYB Introducer, Catheter
Reason for recall
Sterility compromised: Cordis found a pin hole in one of the packaging tray cavities.On January 28, 2010 Cordis Corporation initiated a recall on their AVANTI + Introducer. During routine inspection, a visual irregularity was detected in packing trays.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Cordis initiated the recall January 28, 2010. (The packets were sent overnight on January 28, 2010, with first receipt by customers January 29, 2010.) Immediate corrective action was taken through an inspection and removal of all affected products.
Quantity in commerce
1850 units
Distribution
Nationwide.
Product codes and lots
Catalog 402656X Lot # 15076369.  Catalog 504606X Lot # 15075717.  Catalog 504656X Lot # 15076362, 15076363, and 15076364.