Z-1163-2010

Olympus Trocar Spikes, Model numbers: A5821, A5948, A5949, WA58341L, and WA58343L Used in conjunction with a...

OLYMPUS Corporation of the Americas / initiated 2010-02-08 / Terminated / root cause: manufacturing process

Olympus America Inc. began a recall of Olympus Trocar Spikes, Model numbers: A5821, A5948, A5949, WA58341L, and WA58343L Used in conjunction with a compatible trocar tube, trocar tubes are used for... on . The reason FDA records is "weak weld seams". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1163-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus Trocar Spikes, Model numbers: A5821, A5948, A5949, WA58341L, and WA58343L Used in conjunction with a compatible trocar tube, trocar tubes are used for intraperitoneal access during laparoscopic access during laparoscopic applications, and enable the passage of endoscopes and other diagnostic and therapeutic tools through the abdominal wall.
Product code
KOG
Reason for recall
weak weld seams
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued a Medical Device Recall letter dated February 8, 2010, informing their customers of the problem and the need to return the product.
Quantity in commerce
45
Distribution
Nationwide: AL, AR, CA, FL, MI, MN, MO, NY, PA, OH, TX, and WY. Foreign: Canada
Product codes and lots
Lot numbers: 075W, 076W, 077W, 078W, 081W, 083W, 084W, 07XW, and 07YW