Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0846-2011

The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as...

Neusoft Medical Systems Co., Ltd. / initiated 2010-02-10 / Terminated

Philips And Neusoft Medical Systems Co., Ltd. began a recall of The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as NeuViz Dual, PN: 989605651321. Labeling on the device... on . The reason FDA records is "Potential for the CT patient table top to become detached from the carrier due to missing washers". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0846-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Neusoft Medical Systems Co., Ltd.
Product
The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as NeuViz Dual, PN: 989605651321. Labeling on the device reads in part: "***Neusoft NeuViz Dual MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD. ADDRESS: NEUSOFT PARK, HUN NAN INDUSTRIAL AREA, SHENYANG 110179, CHINA***". Device is intended to produce cross-section images of head and body by computer reconstruction of x-ray transmission data taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Potential for the CT patient table top to become detached from the carrier due to missing washers.
FDA root cause
Other
Action
Neusoft Medical Systems Co., China sent an URGENT DEVICE CORRECTION letter dated Feb. 10, 2010 to their distributor in NC. On 02/12/10, The distributor sent the same letter to their consignees. The firm informed this distributor that there was report of the four screws that secure the patient table top to the carrier that did not have the proper washers in place which caused the table top to detach. If the issue would reoccur, there is a potential for the table top to fall and cause patient injury. Customers were instructed to check the patient table top fixing. If it is found that the the assembly is improper customers were instructed to top using the product and contact local Service personnel for repair. Any customer can contact Service Support Department, Neusoft Medical Systems Co, LTD at [email protected] or Service Support Department at [email protected] for questions about this recall.
Quantity in commerce
26 devices
Distribution
Nationwide Distribution including NC
Product codes and lots
Part number: NeuViz Dual 989605651321.  SERIAL NUMBERS:  400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR080031, NDHR090003, and NDHR090012.