Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0...

Varian Medical Systems, Inc. / initiated 2009-12-09 / Terminated / root cause: software

Varian Medical Systems Oncology Systems began a recall of Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds... on . The reason FDA records is "Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1412-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds 8.2.x). For radiation therapy.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan.
Root cause tag
software
FDA root cause
Software design
Action
Recall initiated on 12/09/2009. Varian issued a notification letter to its consignees, including a description and corrective action step. The firm is in the process of preparing a software fix for the issue.
Quantity in commerce
3177 unit
Distribution
Nationwide distribution and Mexico.
Product codes and lots
v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds 8.2.x), with serial numbers starting with H48.