Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0...
Varian Medical Systems Oncology Systems began a recall of Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds... on . The reason FDA records is "Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1412-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds 8.2.x). For radiation therapy.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Recall initiated on 12/09/2009. Varian issued a notification letter to its consignees, including a description and corrective action step. The firm is in the process of preparing a software fix for the issue.
- Quantity in commerce
- 3177 unit
- Distribution
- Nationwide distribution and Mexico.
- Product codes and lots
- v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds 8.2.x), with serial numbers starting with H48.