Z-0848-2010
BuckyDiagnost. General Radiography and Tomography examination of patients in supine, seated, or standing...
Philips North America LLC / initiated 2010-01-21 / Terminated / root cause: labeling and UDI
Philips Medical Systems North America Co. Phillips began a recall of BuckyDiagnost. General Radiography and Tomography examination of patients in supine, seated, or standing positions on . The reason FDA records is "Device failed to display appropriate labels indicating Radiation Symbol along with the required wording". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0848-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- BuckyDiagnost. General Radiography and Tomography examination of patients in supine, seated, or standing positions.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Device failed to display appropriate labels indicating Radiation Symbol along with the required wording.
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Urgent - Field Safety Notice letters, dated January 20 & 21, 2010, were sent to customers. The letter identified the affected products, gave a description of the problem & the hazard involved, stated how to identify affected products, actions to be taken by the customer or user, and actions planned by the firm. Customers are to make sure that the relevant safety precautions are observed and check their operators console (EEC) if the labeling is present as shown in the drawing. If the labeling is missing, a Philips rep should be contacted. For further information or support concerning the issue, a Philips rep should be contacted at 800-722-9377.
- Quantity in commerce
- 16 units
- Distribution
- Devices were distributed through out the US in the following states: AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TX, UT, VA, WA, WI, and WY.
- Product codes and lots
- Site Numbers: 545896, 17673, 62929, 102578, 104781, 250137, 250294, 505841, 538513, 539863, 542610, 545089, 548438, 557949, 43717870, and 44329671.