Z-1215-2010

ADVANCE TIBIAL WEDGE AUGMENT, REF KTAG-W315, SIZE 3/2+, Thickness 15 DEG, SURFACE NON-POROUS, 1 EACH, Wright...

Wright Medical Technology Inc / initiated 2010-02-19 / Terminated / root cause: manufacturing process

Wright Medical Technology Inc began a recall of ADVANCE TIBIAL WEDGE AUGMENT, REF KTAG-W315, SIZE 3/2+, Thickness 15 DEG, SURFACE NON-POROUS, 1 EACH, Wright Medical Technology, 4677 Airline Road, Arlington... on . The reason FDA records is "The product contained screws with the incorrect length (incorrectly contained 15 mm length screws instead of the correct 5 mm length screws)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1215-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wright Medical Technology Inc
Product
ADVANCE TIBIAL WEDGE AUGMENT, REF KTAG-W315, SIZE 3/2+, Thickness 15 DEG, SURFACE NON-POROUS, 1 EACH, Wright Medical Technology, 4677 Airline Road, Arlington, TN 38002. The device is intended for use in total knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions: 1) Noninflammatory degenerative joint disease, 2)inflammatory joint disease, 3) correction of functional deformity, 4) revision procedures where other treatments or devices have failed and 5) treatment of fractures that are unmanageable using other techniques. (For cement use only).
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The product contained screws with the incorrect length (incorrectly contained 15 mm length screws instead of the correct 5 mm length screws).
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm initiated their recall of the product by letter on February 19, 2010. The notification letters were addressed to distributors, hospital administrators, and surgeons with a fax back response form. The letter described the affected product and reason for the recall. Customers were instructed to examine their inventory and return the recalled product to Wright Medical Technology, Inc. Customers may contact Customer Service by calling 1-800-238-7117 for return instructions and replacement inventory. Direct questions regarding this matter to Wright Medical Technology, Inc. by calling 1-800-874-5630.
Quantity in commerce
47 units
Distribution
Worldwide Distribution -- United States, Canada, Netherlands, Germany, Japan, S. Korea, UK, Czech Republic, Turkey and Brazil.
Product codes and lots
Lot numbers: 087441267 and 097441268.