Z-1255-2010
Foundation Knee System, Patella Trial, 26 mm, part number 802-01-019, Lot number 53944551
Encore Medical, LP / initiated 2010-02-22 / Terminated / root cause: manufacturing process
Encore Medical, Lp began a recall of Foundation Knee System, Patella Trial, 26 mm, part number 802-01-019, Lot number 53944551 on . The reason FDA records is "One lot of product was incorrectly manufactured. Pegs are larger than specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1255-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Foundation Knee System, Patella Trial, 26 mm, part number 802-01-019, Lot number 53944551
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- One lot of product was incorrectly manufactured. Pegs are larger than specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Firm notified consignees by e-mail on 2/22/2010.
- Quantity in commerce
- 2 units
- Distribution
- United States (FL and ID).
- Product codes and lots
- Part Number 802-01-019 Lot 53944551
FIRST ARTICLE / WEEKLY