Coherence Dosimetrist, version 2.2
Siemens Medical Solutions USA, Inc began a recall of Coherence Dosimetrist, version 2.2. Intended use: Medical charged-particle radiation therapy system on . The reason FDA records is "A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1834-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Coherence Dosimetrist, version 2.2. Intended use: Medical charged-particle radiation therapy system
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose, for the OAR (Organs at Risk) the user only defines the max. dose. The firms engineering team determined that the software defect was repeatable.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Affected consignees will be notified by the Customer Service Engineer to schedule a software update beginning November 2009.
- Quantity in commerce
- 60 units
- Distribution
- Worldwide distribution: USA including: CA, KY, MI, NJ and TN; and countries of: Canada and Mexico.
- Product codes and lots
- Version 2.2, Part No. 081239953 Version 2.2.23 SPI Patch, Part No. 08164365