Z-1834-2011

Coherence Dosimetrist, version 2.2

Siemens Medical Solutions USA, Inc. / initiated 2009-11-06 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Coherence Dosimetrist, version 2.2. Intended use: Medical charged-particle radiation therapy system on . The reason FDA records is "A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1834-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Coherence Dosimetrist, version 2.2. Intended use: Medical charged-particle radiation therapy system
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose, for the OAR (Organs at Risk) the user only defines the max. dose. The firms engineering team determined that the software defect was repeatable.
Root cause tag
software
FDA root cause
Software design
Action
Affected consignees will be notified by the Customer Service Engineer to schedule a software update beginning November 2009.
Quantity in commerce
60 units
Distribution
Worldwide distribution: USA including: CA, KY, MI, NJ and TN; and countries of: Canada and Mexico.
Product codes and lots
Version 2.2, Part No. 081239953 Version 2.2.23 SPI Patch, Part No. 08164365