Z-2402-2012Class II

A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table

STERIS Corporation / initiated 2010-02-17 / Terminated / root cause: manufacturing process

Steris Corporation began a recall of A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use... on . FDA classified it as Class II and gives the reason as "Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2402-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Steris sent an Urgent Field Correction Notice letter dated February 17, 2010 to their customers. The letter identified the affected product, description of the problem and actions to be taken. Customers were instructed to examine their units for any structural damage. If found they are to discontinue use and arrange for a field representative to visit the facility to assess the damages and arrange for repair of the unit. For information or questions regarding STERIS visit to your facility contact STERIS Field Service Dispatch at 1-800-333-8828.
Quantity in commerce
11,044 Units
Distribution
Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico and the following countries: American Samoa, Argentina, Australia, Bahrain, Bangladesh, Bermuda, Brazil, Canada, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, Germany, Greece, Guam, Hong Kong, India, Indonesia, Israel, Italy, Japa, Jordan, Kazakhstan, Korea, Kuwait, Lebanon , Libya , Malaysia , Mexico, Morocco, New Zealand, Panama, Peru, Philippines , Poland, Portugal, Qatar, Russia, Saudi Arabia , Singapore, South Africa, Spain, Sri Lanka, Taiwan, Thailand, Trinidad and Tobago, Turkey & United Kingdom.
Product codes and lots
B420702032 to 0402110097

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2402-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2402-2012Class IITerminatedVoluntary: Firm initiated