BIOMET MICROFIXATION, PATIENT DISTRACTOR DRIVER 10.87cm, (4.28")(L), RADEL, STAINLESS STEEL, NON-STERILE...
Biomet Microfixation, Inc. began a recall of BIOMET MICROFIXATION, PATIENT DISTRACTOR DRIVER 10.87cm, (4.28")(L), RADEL, STAINLESS STEEL, NON-STERILE PRODUCT. Part Number SP-1918 BIOMET MICROFIXATION 1520... on . The reason FDA records is "The patient Distractor Driver, used in Osteotomy procedures, may not ratchet and advance the distraction device as intended. If the distraction device does not advance, bone consolidation may occur and a second surgery...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1319-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biomet Microfixation
- Product
- BIOMET MICROFIXATION, PATIENT DISTRACTOR DRIVER 10.87cm, (4.28")(L), RADEL, STAINLESS STEEL, NON-STERILE PRODUCT. Part Number SP-1918 BIOMET MICROFIXATION 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA.
- Product code
- HWR Driver, Prosthesis
- Reason for recall
- The patient Distractor Driver, used in Osteotomy procedures, may not ratchet and advance the distraction device as intended. If the distraction device does not advance, bone consolidation may occur and a second surgery may be necessary.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees will be notified by letter on 01/15/2010 be either fax, email or mail.
- Quantity in commerce
- 73
- Distribution
- US ( AZ,CA,DC,FL,GA,ID,IL,IN,MO,NC,TX,UT), Japan, Finland, Africa
- Product codes and lots
- Lots: 459660, 478770, 530640.