OrthoHelix Part #HCS-056-32 3.2MM Cannulated Hex Driver NON STERILE MAT: 465 SS The drivers are used to drive...
Orthohelix Surgical Designs Inc began a recall of OrthoHelix Part #HCS-056-32 3.2MM Cannulated Hex Driver NON STERILE MAT: 465 SS The drivers are used to drive screws in the Mini MaxTorque Screw System, which... on . FDA classified it as Class II and gives the reason as "The tips of the drivers are breaking while being used to drive the screws into patients". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0026-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- OrthoHelix Part #HCS-056-32 3.2MM Cannulated Hex Driver NON STERILE MAT: 465 SS The drivers are used to drive screws in the Mini MaxTorque Screw System, which is intended to be used to stabilize and aid in the fusion of fractures and osteotomies involving small bones of the hand, wrist, afoot, and ankle.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- The tips of the drivers are breaking while being used to drive the screws into patients.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- The firm, OrthoHelix Surgical Designs, Inc., sent an "IMPORTANT RECALL NOTICE: Mini MaxTorque HCS-056-32 & HCS-056-38 Drivers" dated January 29, 2010 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to discontinue the use of these drivers immediately and return them to OrthoHelix Surgical Designs Inc, 1065 Medina Rd., Suite 500, Medina, OH 44256; ATTN: Driver Recall. Should you have any questions about system, call 330-247-1145 or the Director of Regulatory Affairs & Quality Assurance at 330-247-1444.
- Quantity in commerce
- 47 Pieces
- Distribution
- Nationwide distribution: USA including the states of CA, CO, FL, GA, IL, MO, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, VA, WA, and WI.
- Product codes and lots
- HSC-056-32-1067051
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0026-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0027-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY