Z-2387-2010

Dimension Vista Chemistry 2 Calibrator, Catalog number KC120. Used to calibrate Phosphate, Salicylate, and...

Siemens Healthcare Diagnostics Inc. / initiated 2010-02-26 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista Chemistry 2 Calibrator, Catalog number KC120. Used to calibrate Phosphate, Salicylate, and Triglycerides methods on the Dimension Vista system on . The reason FDA records is "There is incorrect barcode information, which can potentially cause the Alert Message "Calibrator Insert Missing" to be displayed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2387-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista Chemistry 2 Calibrator, Catalog number KC120. Used to calibrate Phosphate, Salicylate, and Triglycerides methods on the Dimension Vista system.
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
There is incorrect barcode information, which can potentially cause the Alert Message "Calibrator Insert Missing" to be displayed.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recalling firm issued an Urgent field Safety Notice, dated February 2010, to all affected customers informing them of the problem. Customers have been told to run calibrators from cups when calibrating or request a replacement lot. The notification should be forwarded to that this product was further distributed to. Siemens Technical Solutions Center should be contacted at 800-441-9250, if they are any technical questions.
Quantity in commerce
1310 cartons
Distribution
Worldwide Distribution -- USA, Austria, Germany, Netherlands, Italy, France, Denmark, Portugal, Canada, South Korea, Malaysia, Australia, and New Zealand.
Product codes and lots
Lot number 9GM001 exp 7/2010