Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0893-2011

Hemochron Signature Elite Whole Blood Microcoagulation (IVD) System. The system is intended to be used only...

Accriva Diagnostics, Inc. / initiated 2010-03-31 / Terminated / root cause: labeling and UDI

International Technidyne Corp. began a recall of Hemochron Signature Elite Whole Blood Microcoagulation (IVD) System. The system is intended to be used only with test cuvettes that are available from ITC on . The reason FDA records is "There is a mismatch between the class of the laser barcode scanner and the labeling of the Hemochron Signature Elite: some of the devices were inadvertently equipped with class II laser barcode scanner instead of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0893-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accriva Diagnostics, Inc.
Product
Hemochron Signature Elite Whole Blood Microcoagulation (IVD) System. The system is intended to be used only with test cuvettes that are available from ITC.
Product code
RFZ Upc Scanner Reader (Bar Code Reader)
Reason for recall
There is a mismatch between the class of the laser barcode scanner and the labeling of the Hemochron Signature Elite: some of the devices were inadvertently equipped with class II laser barcode scanner instead of the intended class I barcode scanner.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Urgent - Medical Device Recall - Hemochron Signature Elite letters, dated February 24, 2010, by Federal Express to all direct customers. The letter informed customers of the issue and stated that the problem does not negatively affect the performance of the product and to continue using the product as intended. In addition, customers are to notify the appropriate personnel about the notice, complete and return the attached form, and an ITC representative will contact customers to discuss arrangements to re-label their devices. Customers can contact ITC Technical Support at 800-579-2255 (US) or 732-548-5700 ext 4700 (International) if they have any questions regarding this issue and notice. Customers can also email [email protected] with questions.
Quantity in commerce
1640 units (1303 US, 337 foreign)
Distribution
Worldwide Distribution -- United Arab Emirates, Belgium, Canada, Switzerland, Germany, Denmark, France, United Kingdom, Ireland, Iceland, Italy, Kuwait, Liechtenstein, Netherlands, Oman, Qatar, Saudi Arabia, Sweden, and Singapore.
Product codes and lots
Serial numbers SE2830 to SE2854 (inclusive); SE2880 to SE4789 (inclusive); SE0260, SE0366, SE0567, SE0764, SE0843, SE0970, SE1086, SE1198, SE1208, SE1212, SE1438, SE1456, SE1580, SE1711, SE2029, SE2136, SE2602, and SE2618.