Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1422-2010

Integra, Licox Ref 1P2.P; Brain PMO-Probe Kit Rx only; Contains Products: REF 1P2 Bolt kit for 2 Brain...

Integra LifeSciences Corporation / initiated 2010-02-18 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of Integra, Licox Ref 1P2.P; Brain PMO-Probe Kit Rx only; Contains Products: REF 1P2 Bolt kit for 2 Brain probes, REF CC1/P1 Combined Oxygen & Temperature Probe... on . The reason FDA records is "One probe that had failed a product release test was inadvertently packaged and released into inventory in Integra's distribution center". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1422-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra, Licox Ref 1P2.P; Brain PMO-Probe Kit Rx only; Contains Products: REF 1P2 Bolt kit for 2 Brain probes, REF CC1/P1 Combined Oxygen & Temperature Probe. Contents Sterile and Non-Pyrogenic if package is unopened and undamaged.
Product code
GWM Device, Monitoring, Intracranial Pressure
Reason for recall
One probe that had failed a product release test was inadvertently packaged and released into inventory in Integra's distribution center.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Integra NeuroSpecialist was sent to retrieve the probe on February 18, 2010, followed by a formal letter.
Quantity in commerce
1
Distribution
One hospital in Omaha, NE.
Product codes and lots
Catalog number CC1P1, Serial number SN 776; Lot number 160110A