Z-1197-2010
Exacta-Mix 2400 Valve Set, 24 - Port Valve Assembly, REF: 724, STERILE, 10 per box. Baxa Corporation...
Baxter Corporation Englewood / initiated 2010-02-10 / Terminated / root cause: manufacturing process
Baxa Corporation began a recall of Exacta-Mix 2400 Valve Set, 24 - Port Valve Assembly, REF: 724, STERILE, 10 per box. Baxa Corporation, Englewood CO. Intended as tubing for use in compounding... on . The reason FDA records is "Molding defect in the valve body causes leaking". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1197-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Corporation Englewood
- Product
- Exacta-Mix 2400 Valve Set, 24 - Port Valve Assembly, REF: 724, STERILE, 10 per box. Baxa Corporation, Englewood CO. Intended as tubing for use in compounding, used to connect up to 24 source ingredients.
- Product code
- LHI Set, I.V. Fluid Transfer
- Reason for recall
- Molding defect in the valve body causes leaking.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Baxa Corporation notified customers of the affected product by letter beginning February 10, 2010. Users were instructed to return all affected product to the firm. For further information, contact Baxa Corporation at 1-303-617-2242.
- Quantity in commerce
- 3230 cases; 32300 units (10 units/case)
- Distribution
- Nationwide Distribution -- Including AL, AR, AZ, CA, CO, DC, MN, IL, NY, DE, FL, PA, IN, TN, MS, MO, TX, MI, MA, MT, MO, SC, VA, WV, NJ, OH, CT, MO, NV, LA, WA, WI, GA, OK, NE, NH, NC, MD, KS, UT, IA, KY, HI, OR, SC, ME and SD.
- Product codes and lots
- Model number: 724, Lot Numbers: 740789 through 742707 (sequentially numbered).