Z-1197-2010

Exacta-Mix 2400 Valve Set, 24 - Port Valve Assembly, REF: 724, STERILE, 10 per box. Baxa Corporation...

Baxter Corporation Englewood / initiated 2010-02-10 / Terminated / root cause: manufacturing process

Baxa Corporation began a recall of Exacta-Mix 2400 Valve Set, 24 - Port Valve Assembly, REF: 724, STERILE, 10 per box. Baxa Corporation, Englewood CO. Intended as tubing for use in compounding... on . The reason FDA records is "Molding defect in the valve body causes leaking". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1197-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Corporation Englewood
Product
Exacta-Mix 2400 Valve Set, 24 - Port Valve Assembly, REF: 724, STERILE, 10 per box. Baxa Corporation, Englewood CO. Intended as tubing for use in compounding, used to connect up to 24 source ingredients.
Product code
LHI Set, I.V. Fluid Transfer
Reason for recall
Molding defect in the valve body causes leaking.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Baxa Corporation notified customers of the affected product by letter beginning February 10, 2010. Users were instructed to return all affected product to the firm. For further information, contact Baxa Corporation at 1-303-617-2242.
Quantity in commerce
3230 cases; 32300 units (10 units/case)
Distribution
Nationwide Distribution -- Including AL, AR, AZ, CA, CO, DC, MN, IL, NY, DE, FL, PA, IN, TN, MS, MO, TX, MI, MA, MT, MO, SC, VA, WV, NJ, OH, CT, MO, NV, LA, WA, WI, GA, OK, NE, NH, NC, MD, KS, UT, IA, KY, HI, OR, SC, ME and SD.
Product codes and lots
Model number: 724, Lot Numbers: 740789 through 742707 (sequentially numbered).