Datascope AS3000 Anesthesia System; Mindray, North America. The AS3000 Anesthesia Delivery System is a...
Mindray DS USA, Inc., dba Datascope Patient Monitoring began a recall of Datascope AS3000 Anesthesia System; Mindray, North America. The AS3000 Anesthesia Delivery System is a continuous flow inhalation gas anesthesia system that... on . The reason FDA records is "An issue affecting the caster mount on the AS3000 System have been identified, specifically reports of the caster mount breaking, causing the wheel to fall off and the unit to tip". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1313-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Datascope AS3000 Anesthesia System; Mindray, North America. The AS3000 Anesthesia Delivery System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor, provides for automatic and manual modes of ventilation, and is equipped with a monitoring system for ventilation, inspired and expired gas. The AS3000 is intended for use in operating rooms. It is used with O2, N2O and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine.
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- An issue affecting the caster mount on the AS3000 System have been identified, specifically reports of the caster mount breaking, causing the wheel to fall off and the unit to tip.
- FDA root cause
- Other
- Action
- An "URGENT PRODUCT FIELD CORRECTIVE ACTION" letter dated March 5, 2010 was sent certified mail to all facilities/customers that have the AS3000 units. The letter described the product, problem and corrective Action/remediation by the firm. A Mindray Service Representative will contact the customers to arrange for the caster mounts to be replaced on the AS3000 units in their facilities at no cost. If you have any questions, please contact Karen Maine of the Mindray DS USA, Inc. Marketing Department at 201-995-8612 or [email protected].
- Quantity in commerce
- 372 units in US, 107 internationally.
- Distribution
- Worldwide Distribution: USA and Mexico, Columbia, Turks & Caicos Islands, BWI; Ecuador, Bolivia, Peru, Dominican Republic, Netherlands Antilles, Chile.SA
- Product codes and lots
- Product numbers 0998-00-3024-01.