Z-1231-2010

Leksell Gamma Knife Perfexion. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Indicated...

Elekta Limited / initiated 2007-05-30 / Terminated / root cause: software

Elekta, Inc. began a recall of Leksell Gamma Knife Perfexion. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Indicated for use in the stereotactic irradiation of... on . The reason FDA records is "There has been an issue with "Image Fushing" where low precision calculation caused images to become inaccurate". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1231-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Leksell Gamma Knife Perfexion. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Indicated for use in the stereotactic irradiation of intracranial structures.
Product code
IWB System, Radiation Therapy, Radionuclide
Reason for recall
There has been an issue with "Image Fushing" where low precision calculation caused images to become inaccurate.
Root cause tag
software
FDA root cause
Software design
Action
Elekta, Inc. issued a Field Change Order, Reference Number 0000946 dated March 14, 2008. Consignees were informed of the affected product and instructed on necessary steps to take. For further information, contact Elekta, Inc. at 1-770-300-9725.
Quantity in commerce
18 units
Distribution
Nationwide Distribution -- AR, AZ, CA, FLA, IL, MI, MN, MO, MS, NJ, OH, OR, PA, UT and VA.
Product codes and lots
Serial Numbers: 6004-6006, 6009-6012, 6014-6018, 6020-6023, 6025, 6027 and 6029-6032.