CareSuite" - Critical Care Manager, PACU Manager and Anesthesia Manager Software: CareSuite 5.1 and later
Picis Inc. began a recall of CareSuite" - Critical Care Manager, PACU Manager and Anesthesia Manager Software: CareSuite 5.1 and later on . The reason FDA records is "Under specific timing of conditions and in configuration with 3rd party infusion pumps, an error within the clinical application causes the manual documentation of clinical orders to be changed to an automatic update...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1252-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Picis Clinical Solutions, Inc.
- Product
- CareSuite" - Critical Care Manager, PACU Manager and Anesthesia Manager Software: CareSuite 5.1 and later
- Product code
- NSX
- Reason for recall
- Under specific timing of conditions and in configuration with 3rd party infusion pumps, an error within the clinical application causes the manual documentation of clinical orders to be changed to an automatic update status. At this point, the order documentation becomes frozen at the last dose and/or rate and the user is unable to modify, stop or delete it. This error could impact the fluid bala
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Picis issued letter notification beginning on March 8, 2010 with software correction information. The letter communicates details of the problem, along with recommendations for mitigating the potential to experience this error condition. Use of the product should be discontinued until the update described in the letter is performed. Customers with questions should call the company at 781-557-3000.
- Quantity in commerce
- 23
- Distribution
- NJ Canada, Australia, Austria,Denmark, Finland, Germany Norway,Portugal, Sweden, France,Netherlands, Spain.
- Product codes and lots
- Software version CareSuite 5.1 and later.