Z-1236-2010
Leksell Gamma Knife. Leksell Gamma Knife is a teletherapy device indicated for use in the stereotactic...
Elekta Limited / initiated 2007-10-01 / Terminated
Elekta, Inc. began a recall of Leksell Gamma Knife. Leksell Gamma Knife is a teletherapy device indicated for use in the stereotactic irradiation of intracranial structures on . The reason FDA records is "After updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1236-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Leksell Gamma Knife. Leksell Gamma Knife is a teletherapy device indicated for use in the stereotactic irradiation of intracranial structures
- Product code
- IWB System, Radiation Therapy, Radionuclide
- Reason for recall
- After updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator.
- FDA root cause
- Other
- Action
- Field Change Order 0000203 "Replacement of obsolete actuator sleigh on Helmet Changer", was sent to all customers that were affected. Direct questions to Elekta, Inc. by calling 1-770-670-2548.
- Quantity in commerce
- 5 units
- Distribution
- United States (CA, TN, TX and WA) and Taiwan.
- Product codes and lots
- LGK 4203, LGK 4309, LGK 4333, LGK 4343 and LGK 4354.