Z-1209-2010

XiO Radiation Treatment Planning System, Release 4.51.00

Elekta Limited / initiated 2010-02-22 / Terminated / root cause: software

Computerized Medical Systems Inc began a recall of XiO Radiation Treatment Planning System, Release 4.51.00 on . The reason FDA records is "Inconsistent MLC positions, dose, and monitor units". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1209-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
XiO Radiation Treatment Planning System, Release 4.51.00.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Inconsistent MLC positions, dose, and monitor units
Root cause tag
software
FDA root cause
Software design
Action
Elekta contacted customers by phone and mail beginning February 22, 2010, explaining the affected product and providing software update information. For further information, contact Elekta at 1-408-830-8023.
Quantity in commerce
Planning system is loaded 25 in locations
Distribution
Worldwide Distribution -- United States (PA, AZ, CA, MN, OH, WI, NC, FL, TX, NY and WI), Australia, The Netherlands and India.
Product codes and lots
Release 4.51.00