Z-1815-2010

CaviBleach Wipes; a disinfectant medical device, active ingredient Sodium hypochlorite 0.525%; a single 9" x...

Medtrol, Inc. / initiated 2009-12-02 / Terminated / root cause: manufacturing process

Medtrol, Inc. began a recall of CaviBleach Wipes; a disinfectant medical device, active ingredient Sodium hypochlorite 0.525%; a single 9" x 9" towelette, pre-moistened with 0.525% - 0.656%... on . The reason FDA records is "The disinfectant wipes were found out of specifications for the disinfectant activity prior to the expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1815-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtrol, Inc.
Product
CaviBleach Wipes; a disinfectant medical device, active ingredient Sodium hypochlorite 0.525%; a single 9" x 9" towelette, pre-moistened with 0.525% - 0.656% sodium hypochlorite, packaged in a flexible poly foil pouch, 50 wipes per dispenser box, 8 boxes per case; Manufactured for Metrex Research Corporation, 28210 Wick Road, Romulus, MI 48174. General purpose disinfectant wipe for hard, non-porous surfaces, patient care equipment and point of care equipment.
Product code
LRJ Disinfectant, Medical Devices
Reason for recall
The disinfectant wipes were found out of specifications for the disinfectant activity prior to the expiration date.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Medtrol telephoned Metrex on December 2, 2009, informing them that Medtrol was requesting the recall of the affected product. Metrex sent Urgent Medical Device Recall letters dated December 3, 2009 to their accounts, requesting them to examine their inventory for the affected product, and return any found for credit. They also requested their accounts to sub-recall the product to the retail level. Metrex also requested their accounts to complete and return the enclosed reply sheet, acknowledging receipt of the recall letter and indicating the amount of product being returned. For further information, contact Metrex Customer Care Center at 1-800-841-1428.
Quantity in commerce
150 cases
Distribution
Nationwide Distribution -- Including state of Michigan.
Product codes and lots
Reorder No. 10-1155, LotF29, Exp June 2010