Z-1232-2010

Leksell Gamma Knife Perfexion, Product Number: 715000. Radionuclide radiation therapy system. Elekta, Inc...

Elekta Limited / initiated 2009-11-06 / Terminated / root cause: software

Elekta, Inc. began a recall of Leksell Gamma Knife Perfexion, Product Number: 715000. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Intended for stereotactic irradiation... on . The reason FDA records is "There may be a situation where the Frame Adapter might lock the stereotactic Frame in the wrong position". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1232-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Leksell Gamma Knife Perfexion, Product Number: 715000. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.
Product code
IWB System, Radiation Therapy, Radionuclide
Reason for recall
There may be a situation where the Frame Adapter might lock the stereotactic Frame in the wrong position.
Root cause tag
software
FDA root cause
Software design
Action
Elekta, Inc. issued a Field Change Order 200 091 titled "Update Frame Adapter with New Frame Fixation Axle" dated November 6, 2009. Consignees were given instructions to prevent further issues related to the affected device. For further information, contact Elekta, Inc. at 1-770-300-9725.
Quantity in commerce
32 units
Distribution
Nationwide Distribution -- AZ, CA, FL, IL, KY, MI, MN, MS, MO, NY, NC, OH, OR, PA, SC, TX, UT and VA.
Product codes and lots
Serial Numbers: 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6045, 6047, 6048, 6049, 6053, 6054, 6056, 6057, 6059 and 6061.