Z-1254-2010
Qwix 4.3mm Stabilization Screws - 34mm
Integra LifeSciences Corporation / initiated 2010-02-08 / Terminated / root cause: labeling and UDI
Integra LifeSciences Corporation began a recall of Qwix 4.3mm Stabilization Screws - 34mm on . The reason FDA records is "The screws were etched incorrectly as 32mm instead of the correct 34mm length". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1254-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Qwix 4.3mm Stabilization Screws - 34mm
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- The screws were etched incorrectly as 32mm instead of the correct 34mm length.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Consignees were notified via conference calls and/or FedEx mailing. Recall aknowledgement forms were provided with written notification. This requires consignees to review their inventory and to indicate any affected lots . This form requires a signature and must be returned.
- Distribution
- Nationwide distribution - Alabama, Arizona, Colorado, Connecticut, Maryland, Michigan, New England, New Jersey, North Carolina, Ohio, South Carolina, Nebraska, and Utah.
- Product codes and lots
- Part Number: 111434SND Lot Number: E6DO