Z-1216-2010

AccuMax Quantum Complete (Model CU3/CU4), an alternating pressure with turn assist & on demand low air loss...

Encompass Therapeutic Support Systems dba BG North America / initiated 2010-03-05 / Terminated / root cause: manufacturing process

Encompass Therapeutic Support Systems dba BG North America began a recall of AccuMax Quantum Complete (Model CU3/CU4), an alternating pressure with turn assist & on demand low air loss system; Encompass Therapuetic Support Systems... on . The reason FDA records is "Potential out gasing of the battery and short term ignition of the gas within the control unit enclosure. This failure mode includes a risk for fire". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1216-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encompass Therapeutic Support Systems dba BG North America
Product
AccuMax Quantum Complete (Model CU3/CU4), an alternating pressure with turn assist & on demand low air loss system; Encompass Therapuetic Support Systems, Glendale, CA. The device is intended to treat or prevent decubitus ulcers, extensive burns, urinary tract blockage and to aid circulation.
Product code
FNM Mattress, Air Flotation, Alternating Pressure
Reason for recall
Potential out gasing of the battery and short term ignition of the gas within the control unit enclosure. This failure mode includes a risk for fire.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An "URGENT RECALL NOTIFICATION" letter dated March 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers should immediately discontinue use of the control unit, turn off the unit and disconnect from the wall socket and quarantine the unit away from any patient's room. The customers should immediately examine your inventory and quarantine the products listed; identify and notify any facility where they may have further distributed or transferred the listed products; and complete and return the REPLY/FAX-BACK FORM as soon as possible to Fax: (626) 796-7856 or mail to EncompassTSS, 100 E. Corson Street, Ste 310, Pasadena, CA 91103. Please contact EncompassTSS Customer Service at 800-822-8288 if you have any questions regarding this recall notice.
Quantity in commerce
349
Distribution
Nationwide distribution: CA, TX, IN, OR, and MO.
Product codes and lots
Serial No: 6101, 6170 and 6267-6613