Z-2281-2010

ABC PROBE REF 160644 5 mm Handswitching Probe, 44 cm QTY: 1 STERILE CONMED CORPORATION 525 French Rd, Utica...

ConMed Corporation / initiated 2008-12-18 / Terminated / root cause: manufacturing process

ConMed Electrosurgery began a recall of ABC PROBE REF 160644 5 mm Handswitching Probe, 44 cm QTY: 1 STERILE CONMED CORPORATION 525 French Rd, Utica, NY 13502... ASSEMBLED IN MEXICO ABC PROBE REF... on . The reason FDA records is "The internal electrode can protrude in front of ceramic insulator at the tip of the ABC Handpiece used with Conmed Electrosurgical units". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2281-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
ABC PROBE REF 160644 5 mm Handswitching Probe, 44 cm QTY: 1 STERILE CONMED CORPORATION 525 French Rd, Utica, NY 13502... ASSEMBLED IN MEXICO ABC PROBE REF 160636 5 mm Handswitching Probe, 44 cm QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13502 ASSEMBLED IN MEXICO Usage: Electrosurgical handpiece for use as monopolar electrosurgical accessory in conjunction with CONMED electrosurgical units equipped with argon beam coagulation.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
The internal electrode can protrude in front of ceramic insulator at the tip of the ABC Handpiece used with Conmed Electrosurgical units.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
An Urgent Device Recall notification was issued 12/18/2008, which explained the problem/risk associated with the device, the discontinuance of use and the return of the device. If you have any questions, please contact 800-552-0138, ext. 5225.
Quantity in commerce
Dom: 46 cs (460 units); Int'l 23 cs (230 units)
Distribution
Nationwide and International
Product codes and lots
Lot: 0306021 through 0806021