Z-1406-2010
PrecisePLAN 2.10 Treatment Planning System
Elekta Limited / initiated 2006-03-15 / Terminated / root cause: software
Elekta, Inc. began a recall of PrecisePLAN 2.10 Treatment Planning System on . The reason FDA records is "Error in dose rates which were calculated at exactly twice the expected values". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1406-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- PrecisePLAN 2.10 Treatment Planning System
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Error in dose rates which were calculated at exactly twice the expected values.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- As described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. The company is directing questions to the local Electa representative or to their offices in the U.K. at telephone 44 (0)-1293-654200.
- Quantity in commerce
- 40 units
- Distribution
- Worldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland
- Product codes and lots
- 1762685762, 1762688017, 1762688506, 1726288507, 1762676911, 1762688175, 1762689384, 1762688019, 1762688022, 1762688876, 1762689442, 1762688177, 1762690813, 1762686657, 1762689588, 1762689386, 1762684357, 1762686833, 1762684370, 1762678238, 1762689519, 1762684359, 1762690814, 1762688514, 1762578223, 1762678239, 1762684360, 1762684361, 1762684362, 1762684367, 1762683737, 1762682414, 1762688081, 1762686584, 1762686843, 1762686832, 1762688024, 1762688080, 1762686836, 1762684358