Z-1432-2010
Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation...
Elekta Limited / initiated 2007-01-11 / Terminated
Elekta, Inc. began a recall of Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use on . The reason FDA records is "In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1432-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.
- Product code
- JAI Couch, Radiation Therapy, Powered
- Reason for recall
- In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
- FDA root cause
- Other
- Action
- Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.
- Quantity in commerce
- 185 units
- Distribution
- Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serbia, Spain, Sudan, Sweden, Taiwan, Thailand, Turkey & United Kingdom.
- Product codes and lots
- 103101, 105136, 105183, 105223, 105271, 105281, 105292, 105330, 105343, 105370, 105446, 105473, 105511, 105518, 105528, 105532, 105542, 105550, 105568, 151416, 151417, 151420, 151421, 151422, 151423, 151424, 151425, 151426, 151427, 151428, 151429, 151430, 151432, 151433, 151434, 151435, 151436, 151437, 151439, 151440, 151441, 151443, 151444, 151445, 151446, 151447, 151448, 151449, 151450, 151451, 151452, 151453, 151454, 151455, 151456, 151457, 151458, 151459, 151460, 151461, 151462, 151463, 151464, 151466, 151467, 151468, 151469, 151470, 151472, 151473, 151474, 151475, 151476, 151477, 151478, 151479, 151480, 151481, 151483, 151484, 151485, 151486, 151487, 151488, 151489, 151490, 151491, 151492, 151493, 151494, 151495, 151496, 151497, 151499, 151500, 151501, 151502, 151503, 151504, 151505, 151506, 151507, 151508, 151509, 151510, 151511, 151513, 151514, 151515, 151516, 151517, 151518, 151519, 151520, 151522, 151523, 151524, 151525, 151526, 151527, 151528, 151529, 151530, 151531, 151532, 151533, 151534, 151535, 151536, 151537, 151538, 151539, 151540, 151541, 151542, 151543, 151544, 151546, 151547, 151549, 151550, 151551, 151552, 151553, 151554, 151555, 151557, 151558, 151559, 151561, 151565, 151566, 151567, 151568, 151569, 151572, 151573, 151578, 151574, 151575, 151579, 151580, 151581, 151583, 151584, 151585, 151587, 151588, 151589, 151590, 151595, 151596, 151597, 151599, 151600, 151601, 151602, 151603, 151608, 151614, 151618, 151622, 151626, 151627, and 151628.