Z-1547-2010

ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll...

ZOLL Medical Corporation / initiated 2010-03-24 / Terminated / root cause: software

ZOLL Medical Corporation, World Wide Headquarters began a recall of ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll Medical Corporation, Chelmsford, MA. Intended for... on . The reason FDA records is "Device issued Shock Advised message but failed to auto-charge the defibrillator". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1547-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZOLL Medical Corporation
Product
ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll Medical Corporation, Chelmsford, MA. Intended for the purpose of converting ventricular fibrillation to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Device issued Shock Advised message but failed to auto-charge the defibrillator.
Root cause tag
software
FDA root cause
Software design
Action
Zoll Medical Corporation issued an "Urgent Medical Device Correction" notification dated March 26, 2010 via certified mail to consignees who have received affected product. Customers will be contacted and given the option to have a software upgrade kit to update their devices or return devices to the firm to have the device upgraded. For further information, contact Zoll Medical Corporation 24/7 Technical Support at 1-800-348-9011 or 1-978-421-9460.
Quantity in commerce
1383 units
Distribution
Worldwide Distribution -- United States, Australia, Germany, Libyan Arab Jamahiriya, Malaysia, Singapore, United Arab Emirates and the United Kingdom.
Product codes and lots
Software versions 4.XX, 6.XX and 7.0X in the E Series BLS Model.