Z-1547-2010
ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll...
ZOLL Medical Corporation / initiated 2010-03-24 / Terminated / root cause: software
ZOLL Medical Corporation, World Wide Headquarters began a recall of ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll Medical Corporation, Chelmsford, MA. Intended for... on . The reason FDA records is "Device issued Shock Advised message but failed to auto-charge the defibrillator". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1547-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Medical Corporation
- Product
- ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll Medical Corporation, Chelmsford, MA. Intended for the purpose of converting ventricular fibrillation to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Device issued Shock Advised message but failed to auto-charge the defibrillator.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Zoll Medical Corporation issued an "Urgent Medical Device Correction" notification dated March 26, 2010 via certified mail to consignees who have received affected product. Customers will be contacted and given the option to have a software upgrade kit to update their devices or return devices to the firm to have the device upgraded. For further information, contact Zoll Medical Corporation 24/7 Technical Support at 1-800-348-9011 or 1-978-421-9460.
- Quantity in commerce
- 1383 units
- Distribution
- Worldwide Distribution -- United States, Australia, Germany, Libyan Arab Jamahiriya, Malaysia, Singapore, United Arab Emirates and the United Kingdom.
- Product codes and lots
- Software versions 4.XX, 6.XX and 7.0X in the E Series BLS Model.