PM1000 Flow Selector - An accessory to an emergency ventilator (resuscitator). The selector is attached to a...
Precision Medical, Inc. began a recall of PM1000 Flow Selector - An accessory to an emergency ventilator (resuscitator). The selector is attached to a flowmeter and is used to direct the outlet flow... on . The reason FDA records is "The on/off outlet is mislabeled so that when a user/operator turns selector to "on", it actually may be turned "off"". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1561-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Precision Medical, Inc.
- Product
- PM1000 Flow Selector - An accessory to an emergency ventilator (resuscitator). The selector is attached to a flowmeter and is used to direct the outlet flow from any standard flowmeter to any one of the three outlets it provides.
- Product code
- BTM Ventilator, Emergency, Manual (Resuscitator)
- Reason for recall
- The on/off outlet is mislabeled so that when a user/operator turns selector to "on", it actually may be turned "off"
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- The recalling firm issued Customer Alert Product Recall letters on 3/5/10 to inform users of the problem and how to identify affected product.
- Quantity in commerce
- 19,518
- Distribution
- The products were shipped to distributor and medical facilities nationwide. The products were also shipped to Taiwan, Germany, Hong Kong, Canada
- Product codes and lots
- Lot numbers 1209 and 0110