Alaris Pump Module model 8100. Subsequent
CareFusion 303, Inc. began a recall of Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids... on . FDA classified it as Class I and gives the reason as "The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2185-2012
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA),subcutaneous, epidural, enteral, or irrigation of fluid spaces. The Pump module is indicated for use on adults, pediatrics, and neonates.
- Product code
- FRN Pump, Infusion
- Reason for recall
- The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- CareFusion sent an Urgent Medical Device Recall Notification letter dated July 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Required action for users instructed high risk patients undergoing infusions at high rates, consider having additional devices as back up. Clinicians should weigh the risk/benefit to the patients before continuing to use the device. Customers with recall related questions were instructed to contact Carefusion support center at (888) 562-6018. Customers with adverse event report questions were instructed to contact customer advocacy at (800) 854-7128, option 1, option 1, option 3 or email at [email protected]. Customers with technical questions regarding the Alaris system were instructed to contact Technical support at (888) 812-3229. For questions regarding this recall call 858-617-4000.
- Quantity in commerce
- 102,248 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: South Africa and Canada. Additional Countries: Netherlands.
- Product codes and lots
- All serial numbers associated with the Alaris Pump Module Model 8100: 8100 DEMO PUMP MODULE; 8100 Global Pump Module; 8100 IOC Pump Module V8.5.29.0; 8100 LVP Training Unit; 8100 Pump Module V8.5.29.0; 8100 Pump Module V8.5.29.0 (V9); Medley Patient Care System LVP; Model 8100 Trade-In Material; RCND 8100 Pump Module V8.5.29.0