Zimmer Reconstruction System Tenaculum 9 inch Length, One 3.5 mm Screw Tine, Item # 00-1179-024-00, Zimmer...
Zimmer Inc. began a recall of Zimmer Reconstruction System Tenaculum 9 inch Length, One 3.5 mm Screw Tine, Item # 00-1179-024-00, Zimmer Inc., Warsaw, IN. Intended to temporarily reduce... on . The reason FDA records is "The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, increasing the potential for fracture". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0496-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Reconstruction System Tenaculum 9 inch Length, One 3.5 mm Screw Tine, Item # 00-1179-024-00, Zimmer Inc., Warsaw, IN. Intended to temporarily reduce pelvic fractures for fixation during reconstructive surgery.
- Product code
- HWN Instrument, Compression
- Reason for recall
- The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, increasing the potential for fracture.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Zimmer sent an Urgent Device Recall letter dated February 16, 2010, to all affected customers. The letter identified the product, the problem, and the action the customers should take. Customers were instructed to monitor all potentially affected patients through regular office consultations. Ensure Zimmer Reconstruction System Tenaculum System Instrument Users at their facility were informed of the urgent device recall. Locate all instruments and quarantine them immediately. Count all affected product in their territory and record the data on the Inventory Return Certification Form included with the letter. Fax a copy of the completed Inventory Return Form to: Zimmer, Inc. at (574) 372-4265. Return the recalled product along with the original completed Inventory Return Certification Form to: Zimmer Distribution Center Attn: Product Service 1777 West Center Street Warsaw, IN 46580 For questions customers were instructed to contact their Zimmer Sales Representative. For questions regarding this recall call 1 - (800) 613-6131.
- Distribution
- Worldwide Distribution - USA including AL, FL, HI, IA, IL, KS KY, LA, MI, MO, MS, NC, ND, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, AND WI and the countries of Australia, Canada, Central or South America, China, Germany, India, Italy, Saudi Arabia, and UK,
- Product codes and lots
- lot #'s: 60636760, 60723667 and 60896770.